Handleiding voor je Philips M3861A HeartStart FR2+ Defibrillator nodig? Hieronder kun je de handleiding gratis bekijken en downloaden als PDF in het Nederlands en/of Engels. Dit product heeft 0 veelgestelde vragen, 0 comments en heeft 0 stemmen. Is dit niet de handleiding die je zoekt, neem dan contact met ons op.

Is je product kapot en biedt de handleiding geen oplossing meer? Dan kun je voor reparatie gratis terecht bij Repair Café.

Handleiding

Loading…

M3860A, M3861A
Edition 15
eÉ~êípí~êí=coOH=aÉÑáÄêáää~íçê
INSTRUCTIONS FOR USE
PHILIPS MEDICAL SYSTEMS
qÜÉ=eÉ~êípí~êí=coOH=aÉÑáÄêáää~íçê
Battery. Standard long-life or
rechargeable battery used to
power the FR2+.
On/Off button. Turns on the
FR2+ and starts voice and text
prompts. Second press turns off
the FR2+.
Status Indicator. Shows you
the readiness status of the FR2+.
Display screen. Displays text
prompts and incident data. The
FR2+ M3860A screen also displays
the patient’s ECG.
Option buttons. Adjust the
contrast of the screen display and
control special functions.
Beeper port. Broadcasts
alert beeps when required. It is
located under the right edge of the
FR2+.
Infrared (IR) communica-
tions port.
A special lens, or
“eye,” used to transfer data
directly to or from another device.
Data card port. Receptacle
for data card tray.
Data card (optional). Used
to store and review information
about an incident, including ECG
and optional voice recording.
Data card tray. Special sleeve
that holds the data card and fits
into the data card port to help seal
the FR2+ against fluids. The tray
should be kept installed in the FR2+
even if no data card is used.
Microphone. Used optionally
to record surrounding audio
during an incident. It is located
under the right edge of the FR2+.
Shock button. Controls
shock delivery. The button flashes
when the HeartStart FR2+ is ready
to deliver a shock.
Speaker. Amplifies voice
prompts during use of the FR2+.
Pads placement diagram.
Illustrates correct placement of
adult pads. Diagrams are also shown
on the defibrillator pads.
Defibrillator pads
connector socket.
Receptacle for
connector of the defibrillator pads
cable. An adjacent LED light flashes
to show socket location and is
covered when connector is
inserted.
Defibrillator pads. Self-
adhesive pads with attached cable
and connector. (Adult pads
shown.)
A
B
C
D
E
F
G
H
I
J
K
L
M
N
O
P
PHILIPS MEDICAL SYSTEMS
eÉ~êípí~êí=coOH=aÉÑáÄêáää~íçê
nrf`h=obcbobk`b=`^oa
PHILIPS MEDICAL SYSTEMS
Intentionally blank.
3-4
PHILIPS MEDICAL SYSTEMS
HEARTSTART FR2+ M3860A/M3861A
Step 3: Shock delivery
First, make sure that no one is touching the patient or the pads. While the FR2+ is
charging, it continues to analyze the patient’s heart rhythm. If the rhythm changes
and a shock is no longer appropriate, the FR2+ disarms. Voice and text prompts
advise you what action to take.
There are four ways you can tell that the defibrillator is ready to deliver a shock:
you hear a voice prompt telling you to deliver a shock,
you see the Shock button flashing,
you hear a steady tone, and/or
you see a text prompt telling you to press the orange (Shock) button.
Press the Shock button to deliver the shock.
IMPORTANT: You must press the button for a shock to be
delivered. The HeartStart FR2+ Defibrillator will not automatically
deliver a shock. This safety feature allows you to verify that the
patient is clear before a shock is delivered.
NOTE:If you do not press the Shock button within 30 seconds of being
prompted, the HeartStart FR2+ will disarm itself and provide a pause. It will
resume analyzing heart rhythm after 30 seconds or when the Resume
Analyzing key is pressed.
After you press the Shock button, a voice prompt tells you the shock was
delivered. Then the FR2+ pauses to allow you to perform CPR. The duration of
this CPR interval is determined by the CPR Timer setting.
CPR Interval
After telling you that it has paused, the FR2+ gives no more voice prompts during
the CPR interval, so that you can provide uninterrupted patient care. During the
CPR interval, the FR2+ screen shows a bar that fills in as the CPR interval is used
up. The HeartStart FR2+ M3860A also displays the ECG, if enabled, during this
period.
NOTE:It is important to perform CPR for the entire duration of the CPR
interval, until you hear the voice prompt telling you to stop CPR.
3-5
PHILIPS MEDICAL SYSTEMS
3
ECG display for ongoing observation
At the discretion of emergency care personnel, the HeartStart FR2+ M3860A with
ECG display enabled can also be used with the M3873A/M3874A FR2+ ECG
assessment module to provide a non-diagnostic ECG display of the patient’s heart
rhythm for attended patient monitoring. The system is intended for use on a
conscious or breathing patient, regardless of age. While connected to the FR2+
ECG assessment module, the FR2+'s shock capability is disabled, but the FR2+
continues to evaluate the patient's ECG. There are no known contraindications to
use of the FR2+ ECG assessment module.
The module is designed for connection to ECG electrodes per AAMI (M3873A) or
IEC (M3874A) color convention. The module’s colored leadwires are connected
to ECG electrodes, which are then placed on the patient’s bare chest, and the
module’s device connector is inserted in the FR2+’s connector socket.
NOTE:It is not necessary to turn the FR2+ Defibrillator off prior to
connecting the ECG assessment module.
Once connected, the HeartStart FR2+ displays and evaluates the patient's ECG
(Lead II). Follow all prompts from the defibrillator. If a data card is used when the
ECG assessment module is connected, all recorded events can be viewed using
one of the Event Review data management software products.
Check the patient if:
indicated by the observed ECG display,
the patient becomes unresponsive or stops breathing, or
the FR2+ prompts
IF NEEDED, ATTACH DEFIBRILLATION PADS.
If appropriate, unplug the ECG assessment module from the FR2+, attach the
defibrillator pads to the patient, and connect the defibrillator pads to the FR2+.
Verify that the defibrillator pads are at least one (1) inch (2.5 cm) away from the
ECG electrodes.
3-6
PHILIPS MEDICAL SYSTEMS
HEARTSTART FR2+ M3860A/M3861A
WARNING:During defibrillation, air pockets between the skin and
defibrillator pads can cause patient skin burns. To help prevent air pockets,
make sure defibrillator pads completely adhere to the skin.
WARNING:Do not let the defibrillator pads touch each other or other ECG
electrodes, lead wires, dressings, transdermal patches, etc. Such contact can
cause electrical arcing and patient skin burns during defibrillation and may
divert defibrillating current away from the heart.
WARNING:Handling or transporting the patient during heart rhythm analysis
can cause an incorrect or delayed diagnosis. If the FR2+ gives a
SHOCK ADVISED
prompt, keep the patient as still as possible for at least 15 seconds so the
HeartStart FR2+ can reconfirm the rhythm analysis before a shock is delivered.
WARNING:CPR rates significantly above 100 compressions per minute can
cause incorrect or delayed analysis by the FR2+. Under certain circumstances,
this may result in a prompt to stop all movement so that the device can
confirm rhythm analysis.
WARNING:Defibrillation current can cause operator or bystander injury. Do
not touch the patient during defibrillation. Do not allow the defibrillator pads
to touch any metal surfaces. Disconnect the pads connector from the FR2+
before using any other defibrillator.
CAUTION: Aggressive handling of the pads in storage or prior to use can
damage the pads. Discard the defibrillator pads if they become damaged.
CAUTION: Do not place the pads directly over an implanted pacemaker or
defibrillator. A noticeable lump with a surgical scar should indicate its position.
PHILIPS MEDICAL SYSTEMS
4-1
4
Q j~áåí~áåáåÖI=qÉëíáåÖI=~åÇ=qêçìÄäÉëÜççíáåÖ
Maintenance
Maintenance of the FR2+ is very simple, but it is a very important factor in its
dependability. When in standby mode (with the battery installed), the FR2+
performs many maintenance activities itself. These include daily and weekly
selftests to verify readiness for use and more detailed monthly selftests that also
verify the shock waveform delivery system. In addition, each time the FR2+ is first
turned on, it executes a selftest. Further, a detailed battery insertion selftest is run
whenever a battery is installed in the FR2+.
The FR2+ requires no calibration or verification of energy delivery. The FR2+ has
no user-serviceable parts.
CAUTION:Improper maintenance may damage the FR2+ or cause it to
function improperly. Maintain the FR2+ only as described in these Instructions
for Use or as designated by your program's Medical Director.
CAUTION:Electrical shock hazard. Dangerous high voltages and currents are
present. Do not open the FR2+, remove its covers, or attempt repair. There
are no user-serviceable components in the FR2+. The FR2+ should be returned
to Philips for repair.
Suggested maintenance schedule
The following table presents a sample maintenance schedule for the FR2+.
Different frequency intervals may be appropriate, depending upon the
environment in which the FR2+ is used. The required maintenance frequency is at
the discretion of your program’s Medical Director.
4-2
PHILIPS MEDICAL SYSTEMS
HEARTSTART FR2+ M3860A/M3861A
After using the HeartStart FR2+
After each use of the FR2+, in addition to the maintenance tasks described in the
table above, perform the following post-use checks before returning the FR2+ to
service:
Check the operation of the FR2+ by removing and reinstalling the battery and
running the battery insertion selftest.
Check the outside of the FR2+ and the connector socket for signs of dirt or
contamination. If the FR2+ is dirty or contaminated, clean it according to the
guidelines provided in this manual.
daily monthly maintenance task/response
Check the Status Indicator.
If you see the flashing black hourglass: The FR2+ is ready to use.
No action required.
If you see anything other than a flashing black hourglass, remove
and reinstall the battery to run the selftest.
If the selftest passes and the Status Indicator shows the
flashing black hourglass, the FR2+ is ready to use.
If the selftest fails, install a new battery and run the selftest. If
the selftest passes, the FR2+ is ready to use. If the selftest
fails, contact Philips Medical Systems.
If the selftest does not run, check to be sure that there is no
pads connector installed in the FR2+.
Check supplies, accessories, and spares for damage and
expiration dating.
Do not use damaged or expired accessories. Replace them
immediately.
If a LOW BATTERY or REPLACE BATTERY message is
displayed: Replace the battery and run the selftest.
Do not attempt to charge the M3863A FR2 standard battery.
It is not rechargeable.
The M3848A FR2+ battery is rechargeable. Recharge it, using
the M3849A Charger only.
Check the outside of the FR2+ and the connector socket
for cracks or other signs of damage.
If you see signs of damage: Contact Philips for technical support.
Download de handleiding in het Nederlands en/of Engels (PDF, 3.96 MB)
(Denk aan het milieu en print deze handleiding alleen als dat echt noodzakelijk is)

Loading…

Meer over deze handleiding

We begrijpen dat het prettig is om een papieren handleiding te hebben voor je Philips M3861A HeartStart FR2+ Defibrillator. Je kunt de handleiding bij ons altijd downloaden en daarna zelf printen. Indien je graag een originele handleiding wilt hebben, raden we je aan contact op te nemen met Philips. Zij kunnen wellicht nog voorzien in een originele handleiding. Zoek je de handleiding van Philips M3861A HeartStart FR2+ Defibrillator in een andere taal? Kies op de homepage je taal naar keuze en zoek daarna op het modelnummer om te zien of we deze beschikbaar hebben.

Specificaties

Merk Philips
Model M3861A HeartStart FR2+
Categorie Defibrillators
Bestandstype PDF
Bestandsgrootte 3.96 MB

Alle handleidingen voor Philips Defibrillators
Meer handleidingen voor Defibrillators

Beoordeling

Laat ons weten wat je van de Philips M3861A HeartStart FR2+ Defibrillator vindt door een beoordeling achter te laten. Wil je je ervaringen met dit product delen of een vraag stellen? Laat dan een bericht achter onderaan de pagina.
Bent u tevreden over de Philips M3861A HeartStart FR2+ Defibrillator?
Ja Nee
Beoordeel dit product als eerste
0 stemmen

Praat mee over dit product

Laat hier weten wat jij vindt van de Philips M3861A HeartStart FR2+ Defibrillator. Als je een vraag hebt, lees dan eerst zorgvuldig de handleiding door. Een handleiding aanvragen kan via ons contactformulier.

Handleiding Philips M3861A HeartStart FR2+ Defibrillator

Gerelateerde producten

Gerelateerde categorieën