CISPR 11 for use in all establishments,including
domestic establishments and those directly
connected to the public low-voltage power
supply network that supplies buildings used
for domestic purposes.
Harmonic
emissions
IEC 61000-3-2
Not applicable
Voltage
fluctuations/
flicker emission
IEC 61000-3-3
Not applicable
Guidance and manufacture’s declaration-electromagnetic immunity
for all EQUIPMENT and SYSTEMS
Guidance and manufacture’s declaration-electromagnetic immunity
The CMS50D1 Pulse Oximeter is intended for use in the electromagnetic environment specified specified
below. The the user of CMS50D1 Pulse Oximeter should assure that it is used in such an environment.
Immunity test IEC60601 test level Compliance level
Electromagnetic environment-
guidance
Electrostatic
discharge (ESD) IEC
61000-4-2
±6KV contact
±8KV air
±6KV contact
±8KV air
Floors should be wood, concrete
or ceramic tile. If floor are
covered with synthetic material,
the relative humidity should be at
least 30%.
Power frequency
(50Hz) magnetic
field
IEC 61000-4-8
3A/m 3A/m Power frequency magnetic fields
should be at levels characteristic
of a typical location in a typical
commercial or hospital
environment
Guidance and manufacture’s declaration-electromagnetic immunity
for EQUIPMENT and SYSTEMS that are not LIFE-SUPPORTING
Guidance and manufacture’s declaration-electromagnetic immunity
The CMS50D1 Pulse Oximeter is intended for use in the electromagnetic environment specified below. The
customer or the user of CMS50D1 Pulse Oximeter should assure that it is used in such an environment.
Immunity
test
IEC60601
test level
Compliance
level
Electromagnetic environment -guidance
Radiated RF
ICE 61000-
4-3
3V/m
80MHz to
2.5GHz
3V/m
Portable and mobile RF communication equipment should be
used no closer to any part of the CMS50D1 Pulse Oximeter,
including cables, than the recommended separation distance
calculated from the equation applicable to the frequency of the
transmitter.
recommended separation distance
80MHz to
800MHz
800MHz to
2.5GHz
Where P is the
maximum output
power rating of the transmitter in watts (W) according to the
transmitter manufacturer and d is the recommended separation
distance in meters (m).
Field strengths from fixed RF transmitters, as determined by an
electromagnetic site survey,
a
should be less than the compliance
level in each frequency range.
b
Interference may occur in the vicinity of equipment marked
with the following symbol:
NOTE 1 At 80MHz and 800MHz, the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by
absorption and reflection from structures, objects and people.
a
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land
mobile radios, amateur radio, AM and FM radio broadcast and TV broadcastcannot be predicted
theoretically with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an
electromagnetic site survey should be considered. If the measured field strength in the location in which
The CMS50D1 Pulse Oximeter is used exceeds the applicable RF compliance level above, the CMS50D1
Pulse Oximeter should be observed to verify normal operation. If abnormal performance is observed,
additional measures may be necessary, such as reorienting or relocating the CMS50D1 Pulse Oximeter.
b
Over the frequency range 150 KHz to 80 MHz, field strengths should be less than 3V/m.
Recommended separation distances between portable and mobile
RF communications equipment and the EQUIPMENT or SYSTEM
for EQUIPMENT or SYSTEM that not LIFE-SUPPORTING
Recommended separation distances between
portable and mobile RF communications equipment and the CMS50D1 Pulse Oximeter
The CMS50D1 Pulse Oximeter is intended for use in the electromagnetic environment in which radiated RF
disturbances are controlled. The customer or the user of the CMS50D1 Pulse Oximeter can help prevent
electromagnetic interference by maintaining a minimum distance between portable and mobile RF
communications equipment (transmitters) and the CMS50D1 Pulse Oximeter as recommended below, according
to the maximum output power of the communications equipment.
Rated maximum output
power of transmitter
(W)
Separation distance according to frequency of transmitter
(m)
150KHz to 80MHz 80MHz to 800MHz
800MHz to 2.5GHz
0.01 0.12 0.12 0.23
0.1 0.37 0.37 0.74
1 1.17 1.17 2.33
10 3.69 3.69 7.38
100 11.67 11.67 23.33
For transmitters rated at a maximum output power not listed above, the recommended separation distanced in
meters (m) can be estimated using the equation applicable to the frequency of the transmitter, where P is the
maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer.
NOTE 1 At 80MHz and 800MHz, the separation distance for the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by
absorption and reflection from structures, objects and people.